FEATURES:
- Transfer rate:
0.1 - 400.0mL / h (10mL syringe).
0.1 - 700.0mL / h (20mL syringe).
0.1 - 900.0mL / h (30mL syringe).
0.1 - 1500.0mL / h (50mL syringe).
Total pump volume range: 0.1 - 999.9mL.
- Transmission accuracy:
+ Regulatory-brand mode
Mechanical accuracy: ± 1%
Pump accuracy: ± 3%
+ Automatic mode select [*] (only 50ml pump can be used).
Mechanical accuracy: ± 1%
Pump accuracy: ± 3%
- Accuracy is achieved when the pump operates for more than 1 hour with a transfer rate of 1.0mL / h or more.
- Bolus speed Approximately 400.0mL / h (10mL syringe).
Approximately 700.0mL / h (20mL syringe).
Approximately 900.0mL / h (30mL syringe).
Approximately 1500.0mL / h (50mL syringe).
- 4-step pressure detecting pressure selected:
Highest:: 119.7 ± 40.0kPa / 900 ± 300mmHg / 1.22 ± 0.41kgf / cm2 ..
Height: 93.1 ± 33.3kPa / 700 ± 250mmHg / 0.95 ± 0.34kgf / cm2.
Average: 66.5 ± 26.7kPa / 500 ± 200mmHg / 0.68 ± 0.27kgf / cm2.
Low: 39.9 ± 20.0kPa / 300 ± 150mmHg / 0.41 ± 0.20kgf / cm2.
- Alarm: Low volume (does not operate in auto-select mode), Obstruction, Low battery, Open syringe clamp, Power switch, Reminder, CPU error.
- Screen number: display 6 digits.
- Power supply: AC power: AC 100 to 240V ± 10%; 50/60 Hz (approx.)
- External DC source: DC 12.7 to 15V (exact)
- Built-in battery: Rechargeable battery, DC12V, 700mAh Ni-Cd.